Overview The i30 current clamp is based on Hall Effect technology for use in measurement of both DC and AC current. The i30 may be used in conjunction with multimeters, recorders and other suitable recording instruments for accurate non-intrusive current measurement. Specifications Electrical Specifications Specified current range 30 mA to 30 A DC, 30 mA to 20 A AC rms Usable current range 5 mA to 30 A DC, 30 mA to 20 A AC rms Crest Factor 1.4 in Output sensitivity 100 mV/A Accuracy (at +25°C) ±1% of reading ±2 mA Resolution ±1 mA Load impedance > 10 k Ohms and ≤ 100pF Conductor position sensitivity ±1% relative to center reading Frequency range DC to 20 kHz (-0.5 dB) Temperature coefficient ±0.01% of reading/°C Power supply 9 V Alkaline, MN1604/PP3, 30 hours, low battery indicator Working voltage (see Safety Standards section) 300 V AC rms or DC General Specifications Maximum conductor size 19 mm diameter Output connection 4 mm safety connector Output zero Manual adjust via thumbwheel Cable length 1.5 meters Operating temperature range 0°C to +50°C Storage temperature range (with battery removed) -20°C to +85°C Operating humidity 15% to 85% (non-condensing) Weight 250 g Safety IEC 61010-1, Pollution Degree 2 IEC 61010-2- 032: CAT III 300V Use of the probe on uninsulated conductors is limited to 300 V ac rms or dc andfrequencies below 1 kHz. EMC Compatibility IEC 61326-1: Portable Electromagnetic Environment Features The i30s current clamp is based on Hall Effect technology for use in measurement of both DC and AC current. The i30s may be used in conjunction with oscilloscopes and other suitable recording instruments for accurate non-intrusive current measurement. I30 Current Probe Data Sheet