Product characteristics Number of inputs and outputs Number of digital outputs: 1; Number of analog outputs: 1 Measuring range -1...100 bar -15...1450 psi -0.1...10 MPa Process connection threaded connection G 1 external thread sealing cone Application System gold-plated contacts Measuring element ceramic-capacitive pressure measuring cell Temperature monitoring no Application flush mountable for the food and beverage industry Media viscous media and liquids with suspended particles; liquids and gases; use in gases at pressures > 25 bar only on request Medium temperature [°C] -25...150 Min. bursting pressure 650 bar 9425 psi 65 MPa Pressure rating 200 bar 2900 psi 20 MPa Vacuum resistance [mbar] -1000 Type of pressure relative pressure; vacuum No dead space yes Electrical data Operating voltage [V] 18...30 DC Min. insulation resistance [MΩ] 100; (500 V DC) Protection class III Reverse polarity protection yes Integrated watchdog yes 2-wire Current consumption [mA] 3.5...21.5 Power-on delay time [s] 1 3-wire Current consumption [mA] < 45 Power-on delay time [s] 0.5 Inputs / outputs Number of inputs and outputs Number of digital outputs: 1; Number of analog outputs: 1 Outputs Total number of outputs 2 Output signal analog signal; IO-Link; (configurable) Number of digital outputs 1; (IO-Link) Number of analog outputs 1 Analog current output [mA] 4...20; (scalable) Max. load [Ω] 700; (Ub = 24 V; (Ub - 9 V) / 21.5 mA) Short-circuit proof yes Overload protection yes Measuring/setting range Measuring range -1...100 bar -15...1450 psi -0.1...10 MPa Analog start point -1...80 bar -15...1160 psi -0.1...8 MPa Analog end point 19...100 bar 276...1450 psi 1.9...10 MPa In steps of 0.05 bar 1 psi 0.005 MPa Factory setting ASP = 0.0 bar AEP = 100.0 bar Accuracy / deviations Repeatability [% of the span] < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1) Characteristics deviation [% of the span] < ± 0,5; (linearity incl. hysteresis and repeatability, limit value setting to DIN EN IEC 62828-1) Linearity deviation [% of the span] < ± 0,15; (Turn down 1:1) Hysteresis deviation [% of the span] < ± 0,15; (Turn down 1:1) Long-term stability [% of the span] < ± 0,1; (Turn down 1:1; per year) Total deviation over temperature range Temperature range total deviation -25...15 °C Characteristics deviation ± 0,15 % of the span / 10 K 15...80 °C Characteristics deviation 80...150 °C Characteristics deviation ± 0,2 % of the span / 10 K Notes on the accuracy / deviation for further details see section Diagrams and graphs Reaction times Damping for the analog output dAA [s] 0...4 2-wire Step response time analog output [ms] 30 3-wire Step response time analog output [ms] 7 Interfaces Communication interface IO-Link Transmission type COM2 (38,4 kBaud) IO-Link revision 1.1 SDCI standard IEC 61131-9 Profiles Digital Measuring Sensor (0x000A), Identification and Diagnosis (0x4000) SIO mode no Required master port class A Process data analog 3 Min. process cycle time [ms]