SPECIFICATIONS: Structure High-purity alumina fiber insulation provides better energy efficiency. Double-layer structure with a cooling fan to keep furnace exteriors safe to touch. One 1“ venting port is built on the top of the furnace One air inlet is located at the back of the furnace. This furnace can be put into a standard glovebox Power Choice of the working voltage 110-120 VAC, single phase, 50/60 Hz, 20 A, 1.8 KVA 208 -240VAC, single phase, 50/60 Hz, 15 A, 2.5 KVA A power cable is included, but without a plug. Operating Temp Range 100-1200°C continuously Control accuracy at 1000 °C +/- 1.0 °C Max. Heating Rate 30°C/ min. Thermocouple K types with a standard plug: 4" Length, 3mm OD. Temperature Controller The digital temperature controller is included. Proportional–integral–derivative control (PID control) and auto-tune function 30 segments programmed with ramping, cooling, and dwelling steps Built-in over-temperature alarm and thermocouple failure alarm +/- 1 ºC temperature control accuracy Default DB9 PC communication port Temperature controller (optional) MTS01 Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Eurotherm temperature controller with +/- 0.1 °C accuracy is available at the extra cost Compliance CE Certified NRTL (UL61010) or CSA certification is available upon request at extra cost. Application Notes For a longer service life, we recommend a heating/cooling rate of less than or equal to 10 °C /min. Small cracks may appear on the surface of the refractory ceramics over time. This is a normal occurrence, and the cracks may be repaired with the alumina patch The heating module is easy to replace. Click the pic left to order a spare Heating module for replacement If firing material releases vapor or smoke, please put the furnace into a fume hood Note Toxic, corrosive, or explosive gases are not allowed for use with this furnace. If you need to open a hole in the top of the furnace ( Pic. 2), please let us know. The furnace can be modified into a hybrid tube furnace (Pic. 3) Pic. 2 Pic. 3