CLEANING AND STERILISATION INSTRUCTIONS 1) Submerge instruments in cleaning agent solution with neutral pH for 20 min; 2) Rinse the instruments thoroughly with purified water for 3-5 min;3) Perform standard moist heat/steam sterilization procedure following the sterilizer instructions WARNING!!!1) Do not use cleaning agents with a high percentage of chlorine and cleaners containing oxalic acid; 2) Do not keep instruments in liquid medium for more than 3 hours Tinesof ring № 1.033 Tinesof ring № 1.099 Three rings listed on this page are compatible with saddle matrices INSTALLATION WITH SPRINGCLIP № 1.003 Choose matrix of appropriate length and insert tines of springclip outside the patient mouth into the matrix side tubes Install system onto the tooth. Operating position on the mesial side.